Drug-induced cellular death dynamics monitored by a highly sensitive organic electrochemical system

Biosens Bioelectron. 2015 Jun 15:68:791-797. doi: 10.1016/j.bios.2015.01.073. Epub 2015 Feb 7.

Abstract

We propose and demonstrate a sensitive diagnostic device based on an Organic Electrochemical Transistor (OECT) for direct in-vitro monitoring cell death. The system efficiently monitors cell death dynamics, being able to detect signals related to specific death mechanisms, namely necrosis or early/late apoptosis, demonstrating a reproducible correlation between the OECT electrical response and the trends of standard cell death assays. The innovative design of the Twell-OECT system has been modeled to better correlate electrical signals with cell death dynamics. To qualify the device, we used a human lung adenocarcinoma cell line (A549) that was cultivated on the micro-porous membrane of a Transwell (Twell) support, and exposed to the anticancer drug doxorubicin. Time-dependent and dose-dependent dynamics of A549 cells exposed to doxorubicin are evaluated by monitoring cell death upon exposure to a range of doses and times that fully covers the protocols used in cancer treatment. The demonstrated ability to directly monitor cell stress and death dynamics upon drug exposure using simple electronic devices and, possibly, achieving selectivity to different cell dynamics is of great interest for several application fields, including toxicology, pharmacology, and therapeutics.

Keywords: Apoptosis/necrosis; Biosensing; Cell death dynamics; Organic Electrochemical Transistor; Pharmacokinetics.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adenocarcinoma / diagnosis
  • Adenocarcinoma of Lung
  • Apoptosis*
  • Biosensing Techniques*
  • Cell Line, Tumor
  • Doxorubicin / pharmacology
  • Electrochemical Techniques*
  • Humans
  • Lung Neoplasms / diagnosis

Substances

  • Doxorubicin