Standard biobanking conditions prevent evaporation of body fluid samples

Clin Chim Acta. 2015 Mar 10:442:141-5. doi: 10.1016/j.cca.2015.01.036. Epub 2015 Feb 4.

Abstract

Pre-analytical variation in biobanking procedures, e.g., long-term storage, could confound biomarker outcomes. We investigated evaporation in various body fluids at different storage temperatures and storage durations. Biobank sample tubes (Sarstedt 72.694.007) filled with water in different volumes (50, 100, 250, 500, 750, 1000, 1250, 1500μl) were stored at different temperatures (-80°C, -20°C, 4°C, room temperature (RT)) for 4.5years and weighed at regular intervals. Next, saliva, serum, plasma, and CSF were stored in different volumes (50, 250, 500, 1000μl) at different temperatures (-80°C, -20°C, 4°C, RT) for 2years. An extra set of CSF was stored in tubes with safe-lock cap (Eppendorf 0030 120.086) instead of a screw cap with o-ring. No evaporation of water stored in biobanking tubes at -80°C or -20°C occurred over 4.5years. Storage of saliva, serum, plasma, and CSF at -80°C or -20°C, monitored over 2years, protected these samples from evaporation too. At 4°C, evaporation was minor, approximately 1.5% (50μl) or 0% (1ml) yearly, where at RT it ranged from 38% (50μl) to 2% (1ml). Differences were observed neither between different body fluids, nor between tube caps. Our data provide support for long-term biobanking conform current biobanking guidelines, encouraging retrospective use of clinical cohorts.

Keywords: Biobanking; Body fluids; Evaporation; Long-term storage.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biological Specimen Banks*
  • Body Fluids / chemistry*
  • Humans
  • Specimen Handling / methods*
  • Temperature
  • Volatilization
  • Water / chemistry

Substances

  • Water