A comparison between conductive and infrared devices for measuring mean skin temperature at rest, during exercise in the heat, and recovery

PLoS One. 2015 Feb 6;10(2):e0117907. doi: 10.1371/journal.pone.0117907. eCollection 2015.

Abstract

Purpose: Skin temperature assessment has historically been undertaken with conductive devices affixed to the skin. With the development of technology, infrared devices are increasingly utilised in the measurement of skin temperature. Therefore, our purpose was to evaluate the agreement between four skin temperature devices at rest, during exercise in the heat, and recovery.

Methods: Mean skin temperature ([Formula: see text]) was assessed in thirty healthy males during 30 min rest (24.0 ± 1.2°C, 56 ± 8%), 30 min cycle in the heat (38.0 ± 0.5°C, 41 ± 2%), and 45 min recovery (24.0 ± 1.3°C, 56 ± 9%). [Formula: see text] was assessed at four sites using two conductive devices (thermistors, iButtons) and two infrared devices (infrared thermometer, infrared camera).

Results: Bland-Altman plots demonstrated mean bias ± limits of agreement between the thermistors and iButtons as follows (rest, exercise, recovery): -0.01 ± 0.04, 0.26 ± 0.85, -0.37 ± 0.98°C; thermistors and infrared thermometer: 0.34 ± 0.44, -0.44 ± 1.23, -1.04 ± 1.75°C; thermistors and infrared camera (rest, recovery): 0.83 ± 0.77, 1.88 ± 1.87°C. Pairwise comparisons of [Formula: see text] found significant differences (p < 0.05) between thermistors and both infrared devices during resting conditions, and significant differences between the thermistors and all other devices tested during exercise in the heat and recovery.

Conclusions: These results indicate poor agreement between conductive and infrared devices at rest, during exercise in the heat, and subsequent recovery. Infrared devices may not be suitable for monitoring [Formula: see text] in the presence of, or following, metabolic and environmental induced heat stress.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Exercise / physiology*
  • Hot Temperature*
  • Humans
  • Infrared Rays*
  • Male
  • Skin Temperature / physiology*
  • Thermometers*