Influence of implant design on blood metal ion concentrations in metal-on-metal total hip replacement patients

Int Orthop. 2015 Sep;39(9):1803-11. doi: 10.1007/s00264-014-2644-z. Epub 2015 Feb 6.

Abstract

Purpose: Most factors considered to affect blood metal ions following metal-on-metal hip replacement are based on hip resurfacing patients. The study aims were to determine which factors affect blood metal ion concentrations following metal-on-metal total hip replacement (MoM THR).

Methods: All unilateral MoM THR patients at one centre with whole-blood cobalt (Co) and chromium (Cr) concentrations measured up to May 2013 were included. Blood sampling was at a mean of 4.5 years (range 1.1-11.8 years) postoperatively.

Results: Of 496 patients (mean age 59.1 years; 52.8% male), blood metal ions >7 μg/l were observed in 9.7% (n = 48). Large femoral head sizes (≥38 mm) had significantly higher (p < 0.0001) blood metal ions than smaller sizes (28/36 mm). Corail-Pinnacle implants produced significantly lower blood metal ions compared to other implant designs (p < 0.01 Co and Cr). Univariate linear regression demonstrated the only significant predictors of both blood Co and Cr concentrations were femoral head size (R(2) = 8.6% Co and R(2) = 3.3% Cr, both p < 0.0001) and implant design (R(2) = 8.8%, p = 0.005 Co and R(2) = 5.1%, p = 0.003 Cr). When the three THR implant design groups (Corail-Pinnacle, Synergy, Other) were analysed separately, femoral head size no longer significantly affected blood metal ions in any of the three implant design groups.

Conclusions: Implant design was the most important factor affecting blood metal ion concentrations. We recommend the regularity of follow-up be tailored to survival rates of various MoM THR designs rather than according to femoral head size.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / adverse effects
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Chromium / blood*
  • Cobalt / blood*
  • Cohort Studies
  • Female
  • Follow-Up Studies
  • Hip Prosthesis / adverse effects*
  • Humans
  • Linear Models
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design / adverse effects*
  • Young Adult

Substances

  • Chromium
  • Cobalt