Scanning electronic microscopy analysis of the apical surface after of root-end resection with different methods

Scanning. 2015 Mar-Apr;37(2):126-30. doi: 10.1002/sca.21188. Epub 2015 Feb 4.

Abstract

To compare the time dispensed and the surface characteristics when a root-end resection was performed with a surgical ultrasonic device (VarioSurg, NSK, Japan) or with a Zekrya bur. Forty distal roots of freshly extracted mandibular molars were selected and divided into four different groups: (G1) stainless steel insert with ultrasound; (G2) diamond insert with ultrasound; (G3) stainless steel and diamond inserts with ultrasound, and (G4) Zekrya burs. The time required for the root-end resection was timed and the surface characteristics were evaluated via scanning electronic microscopy. The median times for the root-end resections were: 115.9 s for G1, 249.8 s for G2, 112.7 s for G3, and 7.44 s for G4, the latter presented a smoother root-end surface. There was no statistically significant difference in the comparisons between G3 group and G4 group (Zekrya burs). The other groups where the root-end resections were with ultrasonic devices proportioned a prolonged time to resection with irregular surfaces.

Keywords: Zekrya bur; root-end resection; ultrasound.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Apicoectomy / methods*
  • Microscopy, Electron, Scanning*
  • Molar / diagnostic imaging
  • Molar / surgery
  • Surface Properties*
  • Time
  • Tooth Apex / diagnostic imaging*
  • Tooth Apex / surgery*
  • Ultrasonography