Differential magnetic force microscope imaging

Scanning. 2015 Mar-Apr;37(2):112-5. doi: 10.1002/sca.21186. Epub 2015 Feb 4.

Abstract

This paper presents a method for differential magnetic force microscope imaging based on a two-pass scanning procedure to extract differential magnetic forces and eliminate or significantly reduce background forces with reversed tip magnetization. In the work, the difference of two scanned images with reversed tip magnetization was used to express the local magnetic forces. The magnetic sample was first scanned with a low lift distance between the MFM tip and the sample surface, and the magnetization direction of the probe was then changed after the first scan to perform the second scan. The differential magnetic force image was obtained through the subtraction of the two images from the two scans. The theoretical and experimental results have shown that the proposed method for differential magnetic force microscope imaging is able to reduce the effect of background or environment interference forces, and offers an improved image contrast and signal to noise ratio (SNR).

Keywords: differential MFM imaging; magnetic force microscope (MFM); reversed tip magnetization.