Examination of wrist and hip actigraphy using a novel sleep estimation procedure ☆

Sleep Sci. 2014 Jun;7(2):74-81. doi: 10.1016/j.slsci.2014.09.007.

Abstract

Objective: Improving and validating sleep scoring algorithms for actigraphs enhances their usefulness in clinical and research applications. The MTI® device (ActiGraph, Pensacola, FL) had not been previously validated for sleep. The aims were to (1) compare the accuracy of sleep metrics obtained via wrist- and hip-mounted MTI® actigraphs with polysomnographic (PSG) recordings in a sample that included both normal sleepers and individuals with presumed sleep disorders; and (2) develop a novel sleep scoring algorithm using spline regression to improve the correspondence between the actigraphs and PSG.

Methods: Original actigraphy data were amplified and their pattern was estimated using a penalized spline. The magnitude of amplification and the spline were estimated by minimizing the difference in sleep efficiency between wrist- (hip-) actigraphs and PSG recordings. Sleep measures using both the original and spline-modified actigraphy data were compared to PSG using the following: mean sleep summary measures; Spearman rank-order correlations of summary measures; percent of minute-by-minute agreement; sensitivity and specificity; and Bland-Altman plots.

Results: The original wrist actigraphy data showed modest correspondence with PSG, and much less correspondence was found between hip actigraphy and PSG. The spline-modified wrist actigraphy produced better approximations of interclass correlations, sensitivity, and mean sleep summary measures relative to PSG than the original wrist actigraphy data. The spline-modified hip actigraphy provided improved correspondence, but sleep measures were still not representative of PSG.

Discussion: The results indicate that with some refinement, the spline regression method has the potential to improve sleep estimates obtained using wrist actigraphy.

Keywords: Actigraphy; Penalized spline; Polysomnography.