A programmable acoustic stimuli and auditory evoked potential measurement system for objective tinnitus diagnosis research

Annu Int Conf IEEE Eng Med Biol Soc. 2014:2014:2749-52. doi: 10.1109/EMBC.2014.6944192.

Abstract

This paper presents the development of a single platform that records auditory evoked potential synchronized to specific acoustic stimuli of the gap prepulse inhibition method for objective tinnitus diagnosis research. The developed system enables to program various parameters of the generated acoustic stimuli. Moreover, only by simple filter modification, the developed system provides high flexibility to record not only short latency auditory brainstem response but also late latency auditory cortical response. The adaptive weighted averaging algorithm to minimize the time required for the experiment is also introduced. The results show that the proposed algorithm can reduce the number of the averaging repetitions to 70% compared with conventional ensemble averaging method.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acoustic Stimulation / methods*
  • Algorithms
  • Biomedical Research
  • Diagnostic Techniques, Otological / instrumentation*
  • Equipment Design
  • Evoked Potentials, Auditory / physiology*
  • Humans
  • Reaction Time
  • Signal Processing, Computer-Assisted / instrumentation*
  • Tinnitus / diagnosis*