Development of a microfluidic-based assay on a novel nitrocellulose platform

Electrophoresis. 2015 Mar;36(6):884-8. doi: 10.1002/elps.201400421. Epub 2015 Feb 20.

Abstract

A novel microfluidic paper-based analytical device (μPAD) utilizing a nitrocellulose (NC) membrane to detect IgG antibodies through a colorimetric analysis is described. The μPAD was constructed using layered polyethylene terephthalate (PET) and pressure-sensitive adhesives (PSA). The biotin labeled Goat Anti-Mouse IgG antibody was spotted and dried on the NC channel prior to subjecting it to a series of wash solutions (Tris-tween), increasing concentrations of alkaline phosphatase conjugated to streptavidin (Strep-ALP), and para-nitrophenyl phosphate (p-NPP) realizing a vibrant yellow color. The reaction proceeds for 10 min before applying the p-NPP stop solution. The device was then dried, scanned, and analyzed yielding a linear range of inverse yellow color intensities versus Strep-ALP concentrations. The development of this simple μPAD should further facilitate the use of NC in colorimetric assays to detect and quantitate antibodies.

Keywords: ELISA; Microfluidic paper-based analytical device; Microfluidics.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Antibodies / chemistry
  • Collodion / chemistry*
  • Colorimetry / instrumentation*
  • Enzyme-Linked Immunosorbent Assay / instrumentation*
  • Equipment Design
  • Mice
  • Microfluidic Analytical Techniques / instrumentation*

Substances

  • Antibodies
  • Collodion