MEMS-based handheld fourier domain Doppler optical coherence tomography for intraoperative microvascular anastomosis imaging

PLoS One. 2014 Dec 4;9(12):e114215. doi: 10.1371/journal.pone.0114215. eCollection 2014.

Abstract

Purpose: To demonstrate the feasibility of a miniature handheld optical coherence tomography (OCT) imager for real time intraoperative vascular patency evaluation in the setting of super-microsurgical vessel anastomosis.

Methods: A novel handheld imager Fourier domain Doppler optical coherence tomography based on a 1.3-µm central wavelength swept source for extravascular imaging was developed. The imager was minimized through the adoption of a 2.4-mm diameter microelectromechanical systems (MEMS) scanning mirror, additionally a 12.7-mm diameter lens system was designed and combined with the MEMS mirror to achieve a small form factor that optimize functionality as a handheld extravascular OCT imager. To evaluate in-vivo applicability, super-microsurgical vessel anastomosis was performed in a mouse femoral vessel cut and repair model employing conventional interrupted suture technique as well as a novel non-suture cuff technique. Vascular anastomosis patency after clinically successful repair was evaluated using the novel handheld OCT imager.

Results: With an adjustable lateral image field of view up to 1.5 mm by 1.5 mm, high-resolution simultaneous structural and flow imaging of the blood vessels were successfully acquired for BALB/C mouse after orthotopic hind limb transplantation using a non-suture cuff technique and BALB/C mouse after femoral artery anastomosis using a suture technique. We experimentally quantify the axial and lateral resolution of the OCT to be 12.6 µm in air and 17.5 µm respectively. The OCT has a sensitivity of 84 dB and sensitivity roll-off of 5.7 dB/mm over an imaging range of 5 mm. Imaging with a frame rate of 36 Hz for an image size of 1000(lateral)×512(axial) pixels using a 50,000 A-lines per second swept source was achieved. Quantitative vessel lumen patency, lumen narrowing and thrombosis analysis were performed based on acquired structure and Doppler images.

Conclusions: A miniature handheld OCT imager that can be used for intraoperative evaluation of microvascular anastomosis was successfully demonstrated.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Arteriovenous Anastomosis / diagnostic imaging*
  • Arteriovenous Anastomosis / physiopathology
  • Diagnostic Imaging*
  • Humans
  • Imaging, Three-Dimensional
  • Mice
  • Micro-Electrical-Mechanical Systems
  • Microvessels / diagnostic imaging*
  • Microvessels / physiopathology
  • Radiography
  • Tomography, Optical Coherence*