The benefits and limitations of reaction cell and sector field inductively coupled plasma mass spectrometry in the detection and quantification of phosphopeptides

Rapid Commun Mass Spectrom. 2015 Jan 15;29(1):35-44. doi: 10.1002/rcm.7079.

Abstract

Rationale: The phosphorylation of proteins is one of the most important post-translational modifications in nature. Knowledge of the quantity or degree of protein phosphorylation in biological samples is extremely important. A combination of liquid chromatography (LC) and inductively coupled plasma mass spectrometry (ICP-MS) allows the absolute and relative quantification of the phosphorus signal.

Methods: A comparison between dynamic reaction cell quadrupole ICP-MS (DRC-Q-ICP-MS) and high-resolution sector field ICP-MS (SF-ICP-MS) in detecting signals of phosphorus-containing species using identical capillary LC (reversed-phase technology) and nebulizer settings was performed.

Results: A method to diminish the reversed-phase gradient-related signal instability in phosphorus detection with LC/ICP-MS applications was developed. Bis(4-nitrophenyl)phosphate (BNPP) was used as a standard to compare signal-to-noise ratios and limits of detection (LODs) between the two instrumental setups. The LOD reaches a value of 0.8 µg L(-1) when applying the DRC technology in Q-ICP-MS and an LOD of 0.09 µg L(-1) was found with the SF-ICP-MS setup. This BNPP standard was further used to compare the absolute quantification possibilities of phosphopeptides in these two setups.

Conclusions: This one-to-one comparison of two interference-reducing ICP-MS instruments demonstrates that absolute quantification of individual LC-separated phosphopeptides is possible. However, based on the LOD values, SF-ICP-MS has a higher sensitivity in detecting phosphorus signals and thus is preferred in phosphopeptide analysis.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid / methods
  • Limit of Detection
  • Mass Spectrometry / methods*
  • Nitrophenols / chemistry
  • Phosphopeptides / analysis*
  • Phosphopeptides / chemistry*
  • Signal-To-Noise Ratio

Substances

  • Nitrophenols
  • Phosphopeptides
  • bis(4-nitrophenyl)phosphate