Validity of physical activity measures in individuals after total knee arthroplasty

Arch Phys Med Rehabil. 2015 Mar;96(3):524-31. doi: 10.1016/j.apmr.2014.10.009. Epub 2014 Oct 31.

Abstract

Objective: To determine the concurrent criterion-related validity of 2 activity monitors in comparison with the criterion method of indirect calorimetry in older adults after total knee arthroplasty (TKA).

Design: Validation study.

Setting: Subjects completed 9 increasingly demanding daily activities in a research laboratory; each activity was performed for 7 minutes, for a total of 80 minutes, while the activity monitors and criterion method were used concurrently.

Participants: Subjects (N=21, 67% women) had a mean age ± SD of 68±7 years and a body mass index of 29±4.

Interventions: Not applicable.

Main outcome measures: Energy expenditure (in kcal/min) measured by accelerometer-based and multisensor-based monitors and by a criterion method. Validity was assessed by the paired t test, intraclass correlation coefficient (ICC), and Bland-Altman plots comparing the measurements from the activity monitors with those of the criterion method.

Results: Measurements from the accelerometer-based monitor were significantly lower than those of the criterion method across all walking and nonwalking activities. The underestimations ranged from 40% to 100%. The accelerometer-based monitor demonstrated small to moderate agreement compared with the criterion method (ICCs from 0 to .38). Measurements from the multisensor-based monitor were significantly lower than those of the criterion method during several nonwalking activities; yet, the differences were minor (2%-19%). Measurements from the multisensor-based monitor during walking activities were not different compared with the criterion method. The multisensor-based monitor demonstrated moderate to excellent agreement with the criterion method (ICCs from .48 to .81).

Conclusions: The multisensor-based monitor showed better criterion-related validity than the accelerometer-based monitor and should be considered as a tool to measure physical activity in individuals after TKA.

Keywords: Accelerometry; Activities of daily living; Arthroplasty; Energy metabolism; Reference standards; Rehabilitation; Validation studies; knee; replacement.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Accelerometry
  • Activities of Daily Living
  • Aged
  • Arthroplasty, Replacement, Knee*
  • Calorimetry, Indirect
  • Energy Metabolism
  • Female
  • Humans
  • Male
  • Monitoring, Ambulatory / instrumentation*
  • Motor Activity*
  • Osteoarthritis, Knee / surgery
  • Surveys and Questionnaires