Complex differential variance algorithm for optical coherence tomography angiography

Biomed Opt Express. 2014 Oct 7;5(11):3822-32. doi: 10.1364/BOE.5.003822. eCollection 2014 Nov 1.

Abstract

We describe a complex differential variance (CDV) algorithm for optical coherence tomography based angiography. The algorithm exploits both the intensity and phase changes of optical coherence tomography (OCT) signals from flowing blood to achieve high vascular contrast, and also intrinsically reject undesirable phase signals originating from small displacement axial bulk tissue motion and instrument synchronization errors. We present this algorithm within a broader discussion of the properties of OCT signal dynamics. The performance of the algorithm is compared against two other existing algorithms using both phantom measurements and in vivo data. We show that the algorithm provides better contrast for a given number of measurements and equivalent spatial averaging.

Keywords: (170.3010) Image reconstruction techniques; (170.4500) Optical coherence tomography.