Effect of probe geometry and optical properties on the sampling depth for diffuse reflectance spectroscopy

J Biomed Opt. 2014;19(10):107002. doi: 10.1117/1.JBO.19.10.107002.

Abstract

The sampling depth of light for diffuse reflectance spectroscopy is analyzed both experimentally and computationally. A Monte Carlo (MC) model was used to investigate the effect of optical properties and probe geometry on sampling depth. MC model estimates of sampling depth show an excellent agreement with experimental measurements over a wide range of optical properties and probe geometries. The MC data are used to define a mathematical expression for sampling depth that is expressed in terms of optical properties and probe geometry parameters.

MeSH terms

  • Anisotropy
  • Models, Theoretical*
  • Monte Carlo Method
  • Optical Imaging / instrumentation*
  • Optical Imaging / methods*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Spectrum Analysis / instrumentation*
  • Spectrum Analysis / methods*