Spatiotemporal phase unwrapping for real-time phase-contrast flow MRI

Magn Reson Med. 2015 Oct;74(4):964-70. doi: 10.1002/mrm.25471. Epub 2014 Oct 10.

Abstract

Purpose: To develop and evaluate a practical phase unwrapping method for real-time phase-contrast flow MRI using temporal and spatial continuity.

Methods: Real-time phase-contrast MRI of through-plane flow was performed using highly undersampled radial FLASH with phase-sensitive reconstructions by regularized nonlinear inversion. Experiments involved flow in a phantom and the human aorta (10 healthy subjects) with and without phase wrapping for velocity encodings of 100 cm·s(-1) and 200 cm·s(-1) . Phase unwrapping was performed for each individual cardiac cycle and restricted to a region of interest automatically propagated to all time frames. The algorithm exploited temporal continuity in forward and backward direction for all pixels with a "continuous" representation of blood throughout the entire cardiac cycle (inner vessel lumen). Phase inconsistencies were corrected by a comparison with values from direct spatial neighbors. The latter approach was also applied to pixels exhibiting a discontinuous signal intensity time course due to movement-induced spatial displacements (peripheral vessel zone).

Results: Phantom and human flow MRI data were successfully unwrapped. When halving the velocity encoding, the velocity-to-noise ratio (VNR) increased by a factor of two.

Conclusion: The proposed phase unwrapping method for real-time flow MRI allows for measurements with reduced velocity encoding and increased VNR.

Keywords: flow MRI; phase unwrapping; phase-contrast MRI; real-time MRI; velocity-encoded MRI.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Computer Simulation
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging