Non-invasive cardiac output monitoring in preterm infants undergoing patent ductus arteriosus ligation: a comparison with echocardiography

Neonatology. 2014;106(4):330-6. doi: 10.1159/000365278. Epub 2014 Oct 1.

Abstract

Background: Non-invasive cardiac output monitoring (NICOM; NICOM™) may be useful in the management of extremely premature preterm infants.

Objectives: To evaluate a new bioreactance-based method of continuous NICOM in preterm infants following patent ductus arteriosus (PDA) ligation.

Methods: Infants underwent three paired NICOM and echocardiography assessments of stroke volume (SV) and left ventricular output (LVO) in the postoperative period: at 1, 6-8, and 16-18 h postoperatively. NICOM- and echocardiography-measured SV and LVO during those periods were compared using Bland-Altman analysis and the intraclass correlation coefficient (ICC).

Results: Twenty-five infants with a median (interquartile range) gestational age and birth weight of 25.0 weeks (24.5-25.9) and 700 g (615-775), respectively, were included. The overall systematic bias (limits of agreement) across all time points between the NICOM and echocardiography SV readings was 39% (8-69) with NICOM consistently underestimating echocardiography values. There was moderate consistency between NICOM and echocardiography SV values (ICC 0.78, p < 0.001). Compared with the 1-hour scans, the 6- to 8- and 16- to 18-hour scans had increased biases of 7.9% (95% CI 2.5-13.2) and 9.7% (95% CI 3.6-15.8), respectively.

Conclusion: Continuous LVO measurement using NICOM was feasible and demonstrated a consistent systematic bias compared with echocardiography in unstable extremely preterm infants without a PDA ligation. NICOM may be used as a trending tool for continuous monitoring in this population, but wide limits of agreement and increasing bias over time suggest it is not interchangeable with echocardiography. © 2014 S. Karger AG, Basel.

Publication types

  • Comparative Study
  • Evaluation Study
  • Observational Study

MeSH terms

  • Birth Weight
  • Cardiac Output*
  • Cardiac Surgical Procedures*
  • Ductus Arteriosus, Patent / diagnosis*
  • Ductus Arteriosus, Patent / diagnostic imaging
  • Ductus Arteriosus, Patent / physiopathology
  • Ductus Arteriosus, Patent / surgery*
  • Equipment Design
  • Feasibility Studies
  • Female
  • Gestational Age
  • Humans
  • Infant, Extremely Low Birth Weight
  • Infant, Newborn
  • Infant, Premature
  • Ligation
  • Male
  • Monitoring, Intraoperative / instrumentation
  • Monitoring, Intraoperative / methods*
  • Ontario
  • Predictive Value of Tests
  • Prospective Studies
  • Reproducibility of Results
  • Ultrasonography
  • Ventricular Function, Left