Scan time reduction in ²³Na-Magnetic Resonance Imaging using the chemical shift imaging sequence: Evaluation of an iterative reconstruction method

Z Med Phys. 2015 Sep;25(3):275-86. doi: 10.1016/j.zemedi.2014.08.007. Epub 2014 Sep 28.

Abstract

Aim: To evaluate potential scan time reduction in (23)Na-Magnetic Resonance Imaging with the chemical shift imaging sequence (CSI) using undersampled data of high-quality datasets, reconstructed with an iterative constrained reconstruction, compared to reduced resolution or reduced signal-to-noise ratio.

Materials and methods: CSI (23)Na-images were retrospectively undersampled and reconstructed with a constrained reconstruction scheme. The results were compared to conventional methods of scan time reduction. The constrained reconstruction scheme used a phase constraint and a finite object support, which was extracted from a spatially registered (1)H-image acquired with a double-tuned coil. The methods were evaluated using numerical simulations, phantom images and in-vivo images of a healthy volunteer and a patient who suffered from cerebral ischemic stroke.

Results: The constrained reconstruction scheme showed improved image quality compared to a decreased number of averages, images with decreased resolution or circular undersampling with weighted averaging for any undersampling factor. Brain images of a stroke patient, which were reconstructed from three-fold undersampled k-space data, resulted in only minor differences from the original image (normalized root means square error < 12%) and an almost identical delineation of the stroke region (mismatch < 6%).

Conclusion: The acquisition of undersampled (23)Na-CSI images enables up to three-fold scan time reduction with improved image quality compared to conventional methods of scan time saving.

Keywords: (23)Na-MRI; (23)Na-Magnetresonanzbildgebung; Compressed Sensing; Unterabtastung; beschränkte Rekonstruktionsverfahren; compressed sensing; constrained reconstruction; undersampling.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Brain / metabolism*
  • Brain / pathology*
  • Feasibility Studies
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Molecular Imaging / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Sodium / pharmacokinetics*
  • Time Factors

Substances

  • Sodium