Locking plate fixation provides superior fixation of humerus split type greater tuberosity fractures than tension bands and double row suture bridges

Clin Biomech (Bristol, Avon). 2014 Nov;29(9):1003-8. doi: 10.1016/j.clinbiomech.2014.08.013. Epub 2014 Aug 30.

Abstract

Background: This paper aims to determine the strongest fixation method for split type greater tuberosity fractures of the proximal humerus by testing and comparing three fixation methods: a tension band with No. 2 wire suture, a double-row suture bridge with suture anchors, and a manually contoured calcaneal locking plate.

Methods: Each method was tested on eight porcine humeri. A osteotomy of the greater tuberosity was performed 50° to the humeral shaft and then fixed according to one of three methods. The humeri were then placed in a testing apparatus and tension was applied along the supraspinatus tendon using a thermoelectric cooling clamp. The load required to produce 3mm and 5mm of displacement, as well as complete failure, was recorded using an axial load cell.

Findings: The average load required to produce 3mm and 5mm of displacement was 658N and 1112N for the locking plate, 199N and 247N for the double row, and 75N and 105N for the tension band. The difference between the three groups was significant (P<0.01). The average load to failure of the locking plate (810N) was significantly stronger than double row (456N) and tension band (279N) (P<0.05). The stiffness of the locking plate (404N/mm) was significantly greater than double row (71N/mm) and tension band (33N/mm) (P<0.01).

Interpretation: Locking plate fixation provides the strongest and stiffest biomechanical fixation for split type greater tuberosity fractures.

Keywords: Biomechanics; Double-row suture-bridge fixation; Greater tuberosity fracture; Locking plate; Tension band.

MeSH terms

  • Animals
  • Biomechanical Phenomena
  • Bone Plates*
  • Bone Screws
  • Bone Wires
  • Disease Models, Animal
  • Equipment Failure Analysis
  • Female
  • Fracture Fixation, Internal / instrumentation*
  • Fracture Fixation, Internal / methods
  • Humans
  • Humerus / surgery
  • Osteotomy / methods
  • Rotator Cuff / surgery
  • Shoulder Fractures / surgery*
  • Stress, Mechanical
  • Suture Anchors
  • Suture Techniques*
  • Swine