Pattern visual evoked potentials elicited by organic electroluminescence screen

Biomed Res Int. 2014:2014:606951. doi: 10.1155/2014/606951. Epub 2014 Aug 14.

Abstract

Purpose: To determine whether organic electroluminescence (OLED) screens can be used as visual stimulators to elicit pattern-reversal visual evoked potentials (p-VEPs).

Method: Checkerboard patterns were generated on a conventional cathode-ray tube (S710, Compaq Computer Co., USA) screen and on an OLED (17 inches, 320 × 230 mm, PVM-1741, Sony, Tokyo, Japan) screen. The time course of the luminance changes of each monitor was measured with a photodiode. The p-VEPs elicited by these two screens were recorded from 15 eyes of 9 healthy volunteers (22.0 ± 0.8 years).

Results: The OLED screen had a constant time delay from the onset of the trigger signal to the start of the luminescence change. The delay during the reversal phase from black to white for the pattern was 1.0 msec on the cathode-ray tube (CRT) screen and 0.5 msec on the OLED screen. No significant differences in the amplitudes of P100 and the implicit times of N75 and P100 were observed in the p-VEPs elicited by the CRT and the OLED screens.

Conclusion: The OLED screen can be used as a visual stimulator to elicit p-VEPs; however the time delay and the specific properties in the luminance change must be taken into account.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cathode Ray Tube
  • Electroretinography*
  • Evoked Potentials, Visual / physiology*
  • Female
  • Humans
  • Luminescence*
  • Pattern Recognition, Visual / physiology*
  • Young Adult