A cost-effective Z-folding controlled liquid handling microfluidic paper analysis device for pathogen detection via ATP quantification

Biosens Bioelectron. 2015 Jan 15:63:379-383. doi: 10.1016/j.bios.2014.07.070. Epub 2014 Aug 2.

Abstract

A cost-effective microfluidic paper analysis device (μPAD) was developed with a special Z-folding design for controlling the fluidic flowing and substrate transportation. This presented μPAD can be easily fabricated through wax printing by using a solid ink printer which deposits wax onto the surface of a chromatographic paper, and then baked on a hotplate by penetrating the molten wax into the paper to create a hydrophobic barrier. After μPAD fabrication, liquid control and substrate transportation can be easily carried out by twice folding the μPAD following Z shape. The Z folding made two separated reagent holding zone connected while the detection reaction occurred with the connection. In this paper, a pathogens detection indicated by ATP quantification was took as a proof-in-principle application of using this presented μPAD, the limit of detection (LOD) was 1 μM for ATP detection and 2.6×10(7) CFU/mL for Salmonella live cell detection, which showed a great potential for Point-of-Care Testing (POCT) applications.

Keywords: ATP detection; Liquid control.; Microfluidic paper analysis device; Pathogen detection.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adenosine Triphosphate / chemistry
  • Adenosine Triphosphate / isolation & purification*
  • Biosensing Techniques / instrumentation*
  • Limit of Detection
  • Microfluidic Analytical Techniques
  • Microfluidics / instrumentation*
  • Printing
  • Salmonella / isolation & purification*
  • Salmonella / pathogenicity

Substances

  • Adenosine Triphosphate