A catheter-based acoustic interrogation device for monitoring motility dynamics of the lower esophageal sphincter

Sensors (Basel). 2014 Aug 12;14(8):14700-11. doi: 10.3390/s140814700.

Abstract

This paper presents novel minimally-invasive, catheter-based acoustic interrogation device for monitoring motility dynamics of the lower esophageal sphincter (LES). A micro-oscillator actively emitting sound wave at 16 kHz is located at one side of the LES, and a miniature microphone is located at the other side of the sphincter to capture the sound generated from the oscillator. Thus, the dynamics of the opening and closing of the LES can be quantitatively assessed. In this paper, experiments are conducted utilizing an LES motility dynamics simulator. The sound strength is captured by the microphone and is correlated to the level of LES opening and closing controlled by the simulator. Measurements from the simulator model show statistically significant (p < 0.05) Pearson correlation coefficients (0.905 on the average in quiet environment and 0.736 on the average in noisy environment, D.O.F. = 9). Measuring the level of LES opening and closing has the potential to become a valuable diagnostic technique for understanding LES dysfunction and the disorders associated with it.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acoustics / instrumentation*
  • Biological Clocks / physiology
  • Catheters
  • Environment
  • Esophageal Sphincter, Lower / physiology*
  • Humans
  • Monitoring, Physiologic / instrumentation*
  • Noise