Full-field fluorescence mode micro-XANES imaging using a unique energy dispersive CCD detector

Anal Chem. 2014 Sep 2;86(17):8791-7. doi: 10.1021/ac502016b. Epub 2014 Aug 19.

Abstract

X-ray absorption near-edge structure (XANES) spectroscopy is a well-known nondestructive technique that allows for chemical state and local structure determination. Spatially resolved oxidation state imaging is possible performing full-field transmission mode XANES experiments, providing chemical state information on the illuminated sample area, but these experiments are limited to thin, concentrated samples. Here we present the use of a unique energy dispersive (ED) pnCCD detector, the SLcam, for full-field fluorescence mode XANES experiments, thereby significantly relaxing the constraints on sample thickness. Using this new detection methodology, spatially resolved chemical state information on millimeter-sized sample areas can be obtained with microscopic resolution in moderate measuring times (less than 15 h), obtaining a XANES profile for each of nearly 70,000 points in a single measurement without the need of scanning the sample through the beam. Besides a description of the use of this detector for micro-XANES applications, we also present the proof of concept for fluorescence mode micro-XANES using a Fe(0)/Fe2O3 model sample and a Nitisol soil sample, which was measured to obtain iron chemical state distribution information.