Comparison of a novel bedside portable endoscopy device with nasogastric aspiration for identifying upper gastrointestinal bleeding

World J Gastroenterol. 2014 Jul 7;20(25):8221-8. doi: 10.3748/wjg.v20.i25.8221.

Abstract

Aim: To compare outcomes using the novel portable endoscopy with that of nasogastric (NG) aspiration in patients with gastrointestinal bleeding.

Methods: Patients who underwent NG aspiration for the evaluation of upper gastrointestinal (UGI) bleeding were eligible for the study. After NG aspiration, we performed the portable endoscopy to identify bleeding evidence in the UGI tract. Then, all patients underwent conventional esophagogastroduodenoscopy as the gold-standard test. The sensitivity, specificity, and accuracy of the portable endoscopy for confirming UGI bleeding were compared with those of NG aspiration.

Results: In total, 129 patients who had GI bleeding signs or symptoms were included in the study (age 64.46 ± 13.79, 91 males). The UGI tract (esophagus, stomach, and duodenum) was the most common site of bleeding (81, 62.8%) and the cause of bleeding was not identified in 12 patients (9.3%). Specificity for identifying UGI bleeding was higher with the portable endoscopy than NG aspiration (85.4% vs 68.8%, P = 0.008) while accuracy was comparable. The accuracy of the portable endoscopy was significantly higher than that of NG in the subgroup analysis of patients with esophageal bleeding (88.2% vs 75%, P = 0.004). Food material could be detected more readily by the portable endoscopy than NG tube aspiration (20.9% vs 9.3%, P = 0.014). No serious adverse effect was observed during the portable endoscopy.

Conclusion: The portable endoscopy was not superior to NG aspiration for confirming UGI bleeding site. However, this novel portable endoscopy device might provide a benefit over NG aspiration in patients with esophageal bleeding.

Keywords: Endoscopy; Gastrointestinal bleeding; Nasogastric aspiration.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Endoscopes, Gastrointestinal*
  • Endoscopy, Gastrointestinal / adverse effects
  • Endoscopy, Gastrointestinal / instrumentation*
  • Equipment Design
  • Female
  • Gastrointestinal Hemorrhage / diagnosis*
  • Gastrointestinal Hemorrhage / etiology
  • Humans
  • Intubation, Gastrointestinal* / adverse effects
  • Male
  • Middle Aged
  • Point-of-Care Systems*
  • Predictive Value of Tests
  • Prospective Studies
  • Suction