Clinical implication of interface pressure for a new prosthetic suspension system

Biomed Eng Online. 2014 Jun 30:13:89. doi: 10.1186/1475-925X-13-89.

Abstract

Background: Prosthesis suspension systems can alter the distribution of pressure within the prosthetic socket. This study evaluates a new suspension system for lower limb prostheses, and aims to compare the interface pressure and amputees' satisfaction with the new system compared with a common prosthetic suspension system (pin/lock).

Methods: Ten transtibial amputees walked at a self-selected speed on a level ground with two different suspension systems, namely the pin/lock and HOLO system. The interface pressure was measured using the F-socket transducers at the proximal, middle and distal sites of residual limb. Furthermore, subjective feedback was logged to compare two systems.

Results: The pressure was significantly higher at the proximal and distal areas with the pin/lock suspension system during the swing phase of gait (P<0.05). Subjective feedback also showed traction at the stump with the pin/lock system. There were no significant differences in the pressure applied to the mid-anterior and mid posterior stump for both suspension systems. However, the lateral and medial sides exhibited higher pressure with the new system during stance phase.

Conclusions: The intention of this study was to deepen understanding on the effect of suspension system on the load distribution over the residual limb. The new coupling system was proved compatible with the pin/lock system in terms of suspending the leg and amputee's satisfaction. On the other hand, the HOLO system could distribute the pressure more uniformly over the residual limb.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Humans
  • Lower Extremity
  • Middle Aged
  • Pressure*
  • Prosthesis Design*
  • Young Adult