[Evaluation of a heart-type fatty acid-binding protein detection kit for diagnosis of acute coronary syndrome]

Nan Fang Yi Ke Da Xue Xue Bao. 2014 Jun;34(6):901-3.
[Article in Chinese]

Abstract

Objective: To evaluation the performance of a heart-type fatty acid-binding protein (H-FABP) ELISA detection kit in the clinical diagnosis of acute coronary syndrome (ACS).

Methods: Plasma or serum samples from 160 suspected ACS patients hospitalized in General Hospital of PLA were examined using Lanzhou H-FABP reagent kit and Holland H-FABP kit. Correlation of the two kits was evaluated and Kappa test was used to examine the consistency of the results of the two products.

Results: The sensitivity of H-FABP diagnosis of ACS detection Lanzhou kit was 91.8%, the specificity was 88.7%, and the total diagnostic rate was 90.42%. The sensitivity of H-FABP diagnosis of ACS detection Holland kit was 90.3%, the specificity was 86.8%, and the total diagnostic rate was 88.75%. The test results showed that two products yielded comparable results (P=0.668, >0.05) with a good consistency (Kappa=0.726, P<0.01).

Conclusions: H-FABP ELISA detection kit produccted by LanZhou biological research institute has a good correlation with H-FABP detection kit produced by HBT company of Holland and has the potential for clinical application.

Publication types

  • Evaluation Study

MeSH terms

  • Acute Coronary Syndrome / diagnosis*
  • Biomarkers / blood
  • Enzyme-Linked Immunosorbent Assay
  • Fatty Acid Binding Protein 3
  • Fatty Acid-Binding Proteins / blood*
  • Humans
  • Reagent Kits, Diagnostic*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Biomarkers
  • FABP3 protein, human
  • Fatty Acid Binding Protein 3
  • Fatty Acid-Binding Proteins
  • Reagent Kits, Diagnostic