Optimization of overlap uniformness for ptychography

Opt Express. 2014 May 19;22(10):12634-44. doi: 10.1364/OE.22.012634.

Abstract

We demonstrate the advantages of imaging with ptychography scans that follow a Fermat spiral trajectory. This scan pattern provides a more uniform coverage and a higher overlap ratio with the same number of scan points over the same area than the presently used mesh and concentric [13] patterns. Under realistically imperfect measurement conditions, numerical simulations show that the quality of the reconstructed image is improved significantly with a Fermat spiral compared with a concentric scan pattern. The result is confirmed by the performance enhancement with experimental data, especially under low-overlap conditions. These results suggest that the Fermat spiral pattern increases the quality of the reconstructed image and tolerance to data with imperfections.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.