Comparison of three plasma sources for ambient desorption/ionization mass spectrometry

J Am Soc Mass Spectrom. 2014 Sep;25(9):1528-37. doi: 10.1007/s13361-014-0924-x. Epub 2014 Jun 4.

Abstract

Plasma-based desorption/ionization sources are an important ionization technique for ambient surface analysis mass spectrometry. In this paper, we compare and contrast three competing plasma based desorption/ionization sources: a radio-frequency (rf) plasma needle, a dielectric barrier plasma jet, and a low-temperature plasma probe. The ambient composition of the three sources and their effectiveness at analyzing a range of pharmaceuticals and polymers were assessed. Results show that the background mass spectrum of each source was dominated by air species, with the rf needle producing a richer ion spectrum consisting mainly of ionized water clusters. It was also seen that each source produced different ion fragments of the analytes under investigation: this is thought to be due to different substrate heating, different ion transport mechanisms, and different electric field orientations. The rf needle was found to fragment the analytes least and as a result it was able to detect larger polymer ions than the other sources.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Ions / chemistry*
  • Plasma Gases / chemistry*
  • Spectrometry, Mass, Electrospray Ionization / instrumentation*
  • Temperature

Substances

  • Ions
  • Plasma Gases