Amplitude-based gated phase-controlled rescanning in carbon-ion scanning beam treatment planning under irregular breathing conditions using lung and liver 4DCTs

J Radiat Res. 2014 Sep;55(5):948-58. doi: 10.1093/jrr/rru032. Epub 2014 May 15.

Abstract

Amplitude-based gating aids treatment planning in scanned particle therapy because it gives better control of uncertainty with the gate window. We have installed an X-ray fluoroscopic imaging system in our treatment room for clinical use with an amplitude-based gating strategy. We evaluated the effects of this gating under realistic organ motion conditions using 4 DCT data of lung and liver tumors. 4 DCT imaging was done for 24 lung and liver patients using the area-detector CT. We calculated the field-specific target volume (FTV) for the gating window, which was defined for a single respiratory cycle. Prescribed doses of 48 Gy relative biological effectiveness (RBE)/fraction/four fields and 45 Gy RBE/two fractions/two fields were delivered to the FTVs for lung and liver treatments, respectively. Dose distributions were calculated for the repeated first respiratory cycle (= planning dose) and the whole respiratory data (= treatment dose). We applied eight phase-controlled rescannings with the amplitude-based gating. For the lung cases, D95 of the treatment dose (= 96.0 ± 1.0%) was almost the same as that of the planning dose (= 96.6 ± 0.9%). D(max)/D(min) of the treatment dose (= 104.5 ± 2.2%/89.4 ± 2.6%) was slightly increased over that of the planning dose (= 102.1 ± 1.0%/89.8 ± 2.5%) due to hot spots. For the liver cases, D95 of the treatment dose (= 97.6 ± 0.5%) was decreased by ∼ 1% when compared with the planning dose (= 98.5 ± 0.4%). D(max)/D(min) of the treatment dose was degraded by 3.0%/0.4% compared with the planning dose. Average treatment times were extended by 46.5 s and 65.9 s from those of the planning dose for lung and liver cases, respectively. As with regular respiratory patterns, amplitude-based gated multiple phase-controlled rescanning preserves target coverage to a moving target under irregular respiratory patterns.

Keywords: 4DCT; carbon ion beam; irregular breathing; liver; lung; scanning beam; treatment planning.

Publication types

  • Clinical Trial

MeSH terms

  • Aged
  • Aged, 80 and over
  • Carbon
  • Female
  • Four-Dimensional Computed Tomography / methods*
  • Heavy Ion Radiotherapy / methods*
  • Humans
  • Ions
  • Liver Neoplasms / diagnostic imaging
  • Liver Neoplasms / radiotherapy*
  • Lung Neoplasms / diagnostic imaging
  • Lung Neoplasms / radiotherapy*
  • Male
  • Middle Aged
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Image-Guided / methods*
  • Reproducibility of Results
  • Respiratory-Gated Imaging Techniques / methods*
  • Sensitivity and Specificity
  • Treatment Outcome

Substances

  • Ions
  • Carbon