Fast separation and determination of free myo-inositol by hydrophilic liquid chromatography

Carbohydr Res. 2014 Jun 4:391:55-60. doi: 10.1016/j.carres.2014.03.010. Epub 2014 Mar 20.

Abstract

A fast liquid chromatography method for separation and determination of myo-inositol is reported. Determination of the biologically important isomer of inositols, myo-inositol, was optimized to avoid overlapping to possible interferents according to European Pharmacopoeia (glycerol, d-mannitol) and saccharose. The method in HILIC mode is extremely selective to other carbohydrates which allows to separate myo-inositol from allo- and d-chiro-inositol with resolution 12.3 and 5.2, resp. and this way it enables to separate myo-inostiol from contingent carbohydrates present in a sample matrix. Retention time of myo-inositol was 12min at 10°C, though higher temperatures (25°C or 40°C) or higher water content in the mobile phase could speed up the separation and determination to four minutes. LOD of the method was 9mg/L at 10°C, and 5mg/L at 25°C, resp.

Keywords: ELSD; HILIC; myo-inositol.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid
  • Hydrophobic and Hydrophilic Interactions
  • Inositol / analogs & derivatives
  • Inositol / analysis*
  • Inositol / chemistry*
  • Time Factors

Substances

  • Inositol