Correction for FDG PET dose extravasations: Monte Carlo validation and quantitative evaluation of patient studies

Med Phys. 2014 May;41(5):052502. doi: 10.1118/1.4870979.

Abstract

Purpose: Current procedure guidelines for whole body [18F]fluoro-2-deoxy-D-glucose (FDG)-positron emission tomography (PET) state that studies with visible dose extravasations should be rejected for quantification protocols. Our work is focused on the development and validation of methods for estimating extravasated doses in order to correct standard uptake value (SUV) values for this effect in clinical routine.

Methods: One thousand three hundred sixty-seven consecutive whole body FDG-PET studies were visually inspected looking for extravasation cases. Two methods for estimating the extravasated dose were proposed and validated in different scenarios using Monte Carlo simulations. All visible extravasations were retrospectively evaluated using a manual ROI based method. In addition, the 50 patients with higher extravasated doses were also evaluated using a threshold-based method.

Results: Simulation studies showed that the proposed methods for estimating extravasated doses allow us to compensate the impact of extravasations on SUV values with an error below 5%. The quantitative evaluation of patient studies revealed that paravenous injection is a relatively frequent effect (18%) with a small fraction of patients presenting considerable extravasations ranging from 1% to a maximum of 22% of the injected dose. A criterion based on the extravasated volume and maximum concentration was established in order to identify this fraction of patients that might be corrected for paravenous injection effect.

Conclusions: The authors propose the use of a manual ROI based method for estimating the effectively administered FDG dose and then correct SUV quantification in those patients fulfilling the proposed criterion.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Computer Simulation
  • Fluorodeoxyglucose F18*
  • Heart / diagnostic imaging
  • Humans
  • Linear Models
  • Male
  • Models, Biological
  • Monte Carlo Method
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation
  • Positron-Emission Tomography / methods*
  • Radiation Dosage*
  • Radiopharmaceuticals*
  • Retrospective Studies
  • Torso / diagnostic imaging

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18