Use of an electrocardiographic screening tool to determine candidacy for a subcutaneous implantable cardioverter-defibrillator

Heart Rhythm. 2014 Aug;11(8):1361-6. doi: 10.1016/j.hrthm.2014.04.025. Epub 2014 Apr 19.

Abstract

Background: An electrocardiographic (ECG) screening test has been developed to identify patients being considered for a totally subcutaneous implantable cardioverter-defibrillator (S-ICD) at risk for T-wave oversensing.

Objective: The purpose of this study was to determine the proportion of potential S-ICD recipients who fail the ECG screening test and to identify predictors of failure.

Methods: Patients who already have an ICD but are not receiving antibradycardia pacing are representative of patients who might be considered for an S-ICD. One hundred such outpatients were enrolled in the study. Surface rhythm strips were recorded along the sensing vectors of the S-ICD system and the screening template applied. Clinical and standard ECG characteristics of patients who failed the test were compared to those who passed.

Results: Patients had the following characteristics: 72% male, age 57 ± 16 years, body mass index 29 ± 6 kg/m(2), left ventricular ejection fraction 43% ± 17%, QRS duration 109 ± 23 ms, QTc interval 447 ± 39 ms, 44% had coronary disease, and 55% had heart failure. Among the 100 patients, 8% failed the screening test. There were no differences in patient clinical characteristics and most standard ECG measurements. However, patients with T-wave inversions in standard ECG leads I, II, and aVF had a 45% chance of failing.

Conclusion: Eight percent of potential S-ICD patients were not eligible for the S-ICD after failing the screening test designed to identify patients susceptible to T-wave oversensing. Patients with T-wave inversions in leads I, II, and aVF on a standard ECG were 23 times more likely to fail. More work is needed in S-ICD sensing algorithms to increase patient eligibility for the S-ICD.

Keywords: Defibrillator screening; Implantable cardioverter-defibrillator; Inappropriate shocks; Subcutaneous implantable cardioverter-defibrillator; T-wave oversensing.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arrhythmias, Cardiac / diagnosis*
  • Arrhythmias, Cardiac / physiopathology
  • Arrhythmias, Cardiac / therapy
  • Death, Sudden, Cardiac / prevention & control*
  • Defibrillators, Implantable*
  • Electrocardiography*
  • Female
  • Follow-Up Studies
  • Heart Conduction System / physiopathology*
  • Humans
  • Male
  • Middle Aged
  • Retrospective Studies
  • Treatment Outcome