Inhibition controls for qualitative real-time PCR assays: are they necessary for all specimen matrices?

J Clin Microbiol. 2014 Jun;52(6):2139-43. doi: 10.1128/JCM.03389-13. Epub 2014 Apr 16.

Abstract

A retrospective analysis of 386,706 specimens representing a variety of matrix types used in qualitative real-time PCR assays determined the overall inhibition rate to be 0.87% when the inhibition control was added preextraction to 5,613 specimens and 0.01% when the inhibition control was added postextraction but preamplification in 381,093 specimens. Inhibition rates of ≤ 1% were found for all specimen matrix types except urine and formalin-fixed, paraffin-embedded tissue.

Publication types

  • Evaluation Study

MeSH terms

  • Humans
  • Molecular Diagnostic Techniques / methods*
  • Molecular Diagnostic Techniques / standards*
  • Real-Time Polymerase Chain Reaction / methods*
  • Real-Time Polymerase Chain Reaction / standards*
  • Reference Standards*