High-quality image acquisition by double exposure overlap in dental cone beam computed tomography

Oral Surg Oral Med Oral Pathol Oral Radiol. 2014 Jun;117(6):760-7. doi: 10.1016/j.oooo.2014.02.024. Epub 2014 Mar 13.

Abstract

Objective: With a double exposure overlapping cone beam computed tomography (CBCT) scan technique, using CBCT acquisition radiation dose, the objective was to obtain apparent density similar to that of multidetector computed tomography (MDCT).

Study design: Factory quality-assurance phantom and water phantom were used for the evaluation of apparent density fidelity of iCAT scans in different modes. Each scan's apparent density was analyzed for identical regions using ImageJ, version 1.42q.

Results: The iCAT Classic extended height acquisition with 4-cm central overlap and reconstruction of 2 groups of 300 projections per rotation for the water and quality-assurance CBCT phantoms resulted in improved apparent density fidelity. This apparent density accuracy was superior to that of iCAT scan at high resolution (600 projections during 1 rotation).

Conclusions: Using double exposure overlapping CBCT scans allows the analysis quality to be comparable with that of MDCT.

MeSH terms

  • Cone-Beam Computed Tomography / methods*
  • Humans
  • Imaging, Three-Dimensional
  • Multidetector Computed Tomography
  • Phantoms, Imaging
  • Radiographic Image Interpretation, Computer-Assisted / methods