Determination of pinocembrin in human plasma by solid-phase extraction and LC/MS/MS: application to pharmacokinetic studies

Biomed Chromatogr. 2014 Dec;28(12):1601-6. doi: 10.1002/bmc.3186. Epub 2014 Apr 14.

Abstract

A sensitive, fast and specific method for the quantitation of pinocembrin in human plasma based on high-performance liquid chromatography-tandem mass spectrometry (LC/MS/MS) was developed and validated. Clonazepam was used as the internal standard (IS). After solid-phase extraction of 500 μL plasma, pinocembrin and the IS were separated on a Luna C8 column using the mobile phase composed of acetonitrile-0.3 mm ammonium acetate solution (65:35, v/v) at a flow rate of 0.25 mL/min in isocratic mode. The detection was performed on a triple quadrupole tandem mass spectrometer by multiple reaction monitoring via an electrospray ionization source in negative mode by AB SCIEX Qtrap 5500. The assay was linear from 1 to 400 ng/mL, with within- and between-run accuracy (relative error) from -1.82 to 0.54%, and within- and between-run precision (CV) below 5.25%. The recovery was above 88% for the analyte at 1, 50 and 300 ng/mL. This analytical method was successful for the determination of pinocembrin in human plasma and applied to a pharmacokinetic study of pinocembrin injection in healthy volunteers after intravenous drip administration.

Keywords: LC/MS/MS; human plasma; method validation; pharmacokinetics; pinocembrin.

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Drug Stability
  • Flavanones / blood*
  • Flavanones / chemistry
  • Flavanones / pharmacokinetics*
  • Humans
  • Linear Models
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Solid Phase Extraction / methods*
  • Tandem Mass Spectrometry / methods*

Substances

  • Flavanones
  • pinocembrin