Calibration of CR-39-based thoron progeny device

Radiat Prot Dosimetry. 2014 Jul;160(1-3):169-72. doi: 10.1093/rpd/ncu072. Epub 2014 Apr 9.

Abstract

Radon isotopes and their progenies have proven significant role in respiratory tumour formation. In most cases, the radiological effect of one of the radon isotopes (thoron) and its progenies has been neglected together with its measurement technique; however, latest surveys proved that thoron's existence is expectable in flats and in workplace in Europe. Detectors based on different track detector measurement technologies have recently spread for measuring thoron progenies; however, the calibration is not yet completely elaborated. This study deals with the calibration of the track detector measurement method suitable for measuring thoron progenies using different devices with measurement techniques capable of measuring several progenies (Pylon AB5 and WLx, Sarad EQF 3220). The calibration factor values related to the thoron progeny monitors, the measurement uncertainty, reproducibility and other parameters were found using the calibration chamber. In the future, the effects of the different parameters (aerosol distribution, etc.) will be determined.