Cementless RM Pressfit Cup: a clinical and radiological study of 91 cases with at least four years follow-up

Orthop Traumatol Surg Res. 2014 Jun;100(4 Suppl):S225-9. doi: 10.1016/j.otsr.2014.03.007. Epub 2014 Apr 3.

Abstract

Cementless metal-back acetabular cups have good long-term results, but some problems have appeared due to the shell's stiffness, modularity and required bearing surfaces. The RM Pressfit Cup is a single-piece polyethylene cementless acetabular cup that is covered by a thin layer of titanium. This allows for bone integration without limitations related to the stiffness of a metal-back shell. There is very little published information about this new, innovative implant design. The purpose of this study was to evaluate the clinical and radiological results from a continuous series of 91 cups (85 patients) with a follow-up of at least 4 years. No patients were lost to follow-up. The Harris Hip Score (HHS) was used to assess the clinical outcome. To assess the radiological outcomes, digital X-rays were used to evaluate the cup position and integration; wear was measured using Livermore's technique. The clinical results were excellent: the mean HHS was 94 and 82% of cases had good or excellent scores. Three of the cups had to be revised because of dislocation brought on by incorrect positioning. X-rays revealed that three implants had shifted during the first 6 weeks, but had stabilized afterwards. Bone integration on X-rays was satisfactory in all cases with no signs of osteolysis. The configuration of the bone trabeculae showed that loads between the implant and peri-acetabular cancellous bone were evenly distributed. The wear of the polyethylene cup-ceramic head bearing was 0.07 mm/year. The results of this series are consistent with recent published studies with the RM Pressfit Cup.

Level of evidence: IV.

Keywords: Cementless acetabular cup; Osseointegration; Total hip arthroplasty.

Publication types

  • Evaluation Study

MeSH terms

  • Acetabulum / diagnostic imaging*
  • Acetabulum / physiology
  • Acetabulum / surgery*
  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Equipment Failure Analysis
  • Female
  • Follow-Up Studies
  • Hip Joint / diagnostic imaging
  • Hip Joint / physiology
  • Hip Joint / surgery
  • Hip Prosthesis*
  • Humans
  • Longitudinal Studies
  • Male
  • Middle Aged
  • Osteoarthritis, Hip / surgery
  • Osteogenesis / physiology
  • Osteonecrosis / surgery
  • Polyethylene*
  • Prosthesis Design*
  • Radiography
  • Titanium*
  • Treatment Outcome

Substances

  • Polyethylene
  • Titanium