Design of a microfluidic system for red blood cell aggregation investigation

J Biomech Eng. 2014 Jun;136(6):064501. doi: 10.1115/1.4027351.

Abstract

The purpose of this paper is to design a microfluidic apparatus capable of providing controlled flow conditions suitable for red blood cell (RBC) aggregation analysis. The linear velocity engendered from the controlled flow provides constant shear rates used to qualitatively analyze RBC aggregates. The design of the apparatus is based on numerical and experimental work. The numerical work consists of 3D numerical simulations performed using a research computational fluid dynamics (CFD) solver, Nek5000, while the experiments are conducted using a microparticle image velocimetry system. A Newtonian model is tested numerically and experimentally, then blood is tested experimentally under several conditions (hematocrit, shear rate, and fluid suspension) to be compared to the simulation results. We find that using a velocity ratio of 4 between the two Newtonian fluids, the layer corresponding to blood expands to fill 35% of the channel thickness where the constant shear rate is achieved. For blood experiments, the velocity profile in the blood layer is approximately linear, resulting in the desired controlled conditions for the study of RBC aggregation under several flow scenarios.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Buffers
  • Computer Simulation
  • Equipment Design
  • Erythrocyte Aggregation*
  • Erythrocytes / cytology*
  • Hydrodynamics
  • Microfluidic Analytical Techniques / instrumentation*
  • Plasma / cytology
  • Swine

Substances

  • Buffers