Guidance for accurate and consistent tissue Doppler velocity measurement: comparison of echocardiographic methods using a simple vendor-independent method for local validation

Eur Heart J Cardiovasc Imaging. 2014 Jul;15(7):817-27. doi: 10.1093/ehjci/jeu040. Epub 2014 Apr 2.

Abstract

Background: Variability has been described between different echo machines and different modalities when measuring tissue velocities. We assessed the consistency of tissue velocity measurements across different modalities and different manufacturers in an in vitro model and in patients. Furthermore, we present freely available software tools to repeat these evaluations.

Methods and results: We constructed a simple setup to generate reproducible motion and used it to compare velocities measured using three echocardiographic modalities: M-mode, speckle tracking, and tissue Doppler, with a straightforward, non-ultrasound, optical gold standard. In the clinical phase, 25 patients underwent M-mode, speckle tracking, and tissue Doppler measurements of s', e', and a' velocities. In vitro, the M-mode and speckle tracking velocities agreed with optical assessment. Of the three possible tissue Doppler measurement conventions (outer, middle, and inner edge) only the middle agreed with optical assessment (discrepancy -0.20 (95% CI -0.44 to 0.03) cm/s, P = 0.11, outer +5.19 (4.65 to 5.73) cm/s, P < 0.0001, inner -6.26 (-6.87 to -5.65) cm/s, P < 0.0001). A similar pattern occurred across all four studied manufacturers. M-mode was therefore chosen as the in vivo gold standard. Clinical measurements of s' velocities by speckle tracking and the middle line of the tissue Doppler showed concordance with M-mode, while the outer line overestimated significantly (+1.27(0.96 to 1.59) cm/s, P < 0.0001) and the inner line underestimated (-1.82 (-2.11 to -1.52) cm/s, P < 0.0001).

Conclusions: Echocardiographic velocity measurements can be more consistent than previously suspected. The statistically modal velocity, found at the centre of the spectral pulsed wave tissue Doppler envelope, most closely represents true tissue velocity. This article includes downloadable, vendor-independent software enabling calibration of echocardiographic machines using a simple, inexpensive in vitro setup.

Keywords: Calibration; Echocardiography; Tissue Doppler; Velocity.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Aged
  • Analysis of Variance
  • Blood Flow Velocity
  • Calibration
  • Echocardiography, Doppler, Pulsed / methods*
  • Echocardiography, Doppler, Pulsed / standards*
  • Female
  • Guidelines as Topic*
  • Humans
  • Image Processing, Computer-Assisted*
  • Laser-Doppler Flowmetry / methods
  • Laser-Doppler Flowmetry / standards*
  • Male
  • Middle Aged
  • Phantoms, Imaging*
  • Reproducibility of Results
  • Sampling Studies
  • Sensitivity and Specificity