Preliminary clinical experience with digital breast tomosynthesis in the visualization of breast microcalcifications

J Clin Imaging Sci. 2013 Dec 31:3:65. doi: 10.4103/2156-7514.124099. eCollection 2013.

Abstract

Objectives: To compare the visualization and image quality of microcalcifications imaged with digital breast tomosynthesis (DBT) versus conventional digital mammography.

Materials and methods: Patients with microcalcifications detected on full field digital mammography (FFDM) recommended for needle core biopsy were enrolled in the study after obtaining patient's consent and institutional review board approval (n = 177 patients, 179 lesions). All had a bilateral combination DBT exam, after undergoing routine digital mammography, prior to biopsy. The study radiologist reviewed the FFDM and DBT images in a non-blinded comparison and assessed the visibility of the microcalcifications with both methods, including image quality and clarity with which the calcifications were seen. Data recorded included patient demographics, lesion size on FFDM, DBT, and surgical excision (when applicable), biopsy, and surgical pathology, if any.

Results: Average lesion size on DBT was 1.5 cm; average lesion size on FFDM was 1.4 cm. The image quality of DBT was assessed as equivalent or superior in 92.2% of cases. In 7.8% of the cases, the FFDM image quality was assessed as equivalent or superior.

Conclusion: In our review, DBT image quality appears to be comparable to or better than conventional FFDM in terms of demonstrating microcalcifications, as shown in 92.2% of cases.

Keywords: Breast imaging; digital breast tomosynthesis; digital mammography; microcalcifications.