New concepts and materials for the manufacturing of MR-compatible guide wires

Biomed Tech (Berl). 2014 Apr 1;59(2):147-51. doi: 10.1515/bmt-2013-0027.

Abstract

This paper shows the development of a new magnetic resonance imaging (MRI)-compatible guide wire made from fiber-reinforced plastics. The basic material of the developed guide wire is manufactured using a specially developed micro-pullwinding technology, which allows the adjustment of tensile, bending, and torsional stiffness independent from each other. Additionally, the micro-pullwinding technology provides the possibility to vary the stiffness along the length of the guide wire in a continuous process. With the possibilities of this technology, the mechanical properties of the guide wire were precisely adjusted for the intended usage in MRI-guided interventions. The performance of the guide wire regarding the mechanical properties was investigated. It could be shown, that the mechanical properties could be changed independently from each other by varying the process parameters. Especially, the torsional stiffness could be significantly improved with only a minor influence on bending and tensile properties. The precise influence of the variation of the winding angle on the mechanical and geometrical properties has to be further investigated. The usability of the guide wire as well as its visibility in MRI was investigated by radiologists. With the micro-pullwinding technology, a continuous manufacturing technique for highly stressable, MRI-safe profiles is available and can be the trigger for a new class of medical devices.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biocompatible Materials / chemical synthesis*
  • Catheters*
  • Elastic Modulus
  • Equipment Design
  • Equipment Failure Analysis
  • Fiducial Markers*
  • Magnetic Resonance Imaging / instrumentation*
  • Materials Testing
  • Tensile Strength

Substances

  • Biocompatible Materials