Hip imaging of avascular necrosis at 7 Tesla compared with 3 Tesla

Skeletal Radiol. 2014 May;43(5):623-32. doi: 10.1007/s00256-014-1818-5. Epub 2014 Feb 5.

Abstract

Objectives: To compare ultra-high field, high-resolution bilateral magnetic resonance imaging (MRI) of the hips at 7 Tesla (T) with 3 T MRI in patients with avascular necrosis (AVN) of the femoral head by subjective image evaluations, contrast measurements, and evaluation of the appearance of imaging abnormalities.

Materials and methods: Thirteen subjects with avascular necrosis treated using advanced core decompression underwent MRI at both 7 T and 3 T. Sequence parameters as well as resolution were kept identical for both field strengths. All MR images (MEDIC, DESS, PD/T2w TSE, T1w TSE, and STIR) were evaluated by two radiologists with regard to subjective image quality, soft tissue contrasts, B1 homogeneity (four-point scale, higher values indicating better image quality) and depiction of imaging abnormalities of the femoral heads (three-point scale, higher values indicating the superiority of 7 T). Contrast ratios of soft tissues were calculated and compared with subjective data.

Results: 7-T imaging of the femoral joints, as well as 3-T imaging, achieved "good" to "very good" quality in all sequences. 7 T showed significantly higher soft tissue contrasts for T2w and MEDIC compared with 3 T (cartilage/fluid: 2.9 vs 2.2 and 3.6 vs 2.6), better detailed resolution for cartilage defects (PDw, T2w, T1w, MEDIC, DESS > 2.5) and better visibility of joint effusions (MEDIC 2.6; PDw/T2w 2.4; DESS 2.2). Image homogeneity compared with 3 T (3.9-4.0 for all sequences) was degraded, especially in TSE sequences at 7 T through signal variations (7 T: 2.1-2.9); to a lesser extent also GRE sequences (7 T: 2.9-3.5). Imaging findings related to untreated or treated AVN were better delineated at 3 T (≤1.8), while joint effusions (2.2-2.6) and cartilage defects (2.5-3.0) were better visualized at 7 T. STIR performed much more poorly at 7 T, generating large contrast variations (1.5).

Conclusions: 7-T hip MRI showed comparable results in hip joint imaging compared with 3 T with slight advantages in contrast detail (cartilage defects) and fluid detection at 7 T when accepting image degradation medially.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Female
  • Femur Head Necrosis / pathology*
  • Hip Joint / pathology*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Sensitivity and Specificity