Meta-analysis of bioabsorbable versus durable polymer drug-eluting stents in 20,005 patients with coronary artery disease: an update

Catheter Cardiovasc Interv. 2014 May 1;83(6):E193-206. doi: 10.1002/ccd.25416. Epub 2014 Feb 10.

Abstract

Objectives: To perform an updated meta-analysis comparing biodegradable polymer drug eluting stents (BP-DES) and durable polymer drug eluting stents (DP-DES).

Background: BP-DES have been suggested to reduce late stent thrombosis (LST) rates as compared to first generation DP-DES. Recently, second generation DP-DES have replaced older DES, but comparison of these stents with BP-DES has not yielded consistent results.

Methods: Medline/Web databases were searched for studies comparing BP-DES and DP-DES, and reporting rates of overall/cardiac mortality, myocardial infarction (MI), LST, target lesion revascularization (TLR) and target vessel revascularization (TVR) and late lumen loss (LLL), with a follow-up ≥6 months.

Results: Twenty studies (20,005 patients) were included in the meta-analysis. Median follow-up time was 1 year. Compared with DP-DES, BP-DES showed lower LLL (in stent: weighted mean difference WMD -0.45 mm, 95% CI -0.66 to -0.24 mm, P = 0.00001; in segment: WMD -0.15 mm, 95% CI = -0.24 to -0.06 mm, P = 0.001) and lower rates of LST (OR 0.51, 95% CI = 0.30 to 0.86, P = 0.01), although they did not improve mortality, MI, TLR, and TVR rates. BP-DES coated with sirolimus or novolimus, in comparison with biolimus or paclitaxel, were associated with reduced LLL (P < 0.0001 for subgroups).

Conclusions: In comparison with DP-DES, BP-DES significantly reduce LLL and LST rates, without clear benefits on harder endpoints. The efficacy of BP-DES in preserving lumen patency seems larger for sirolimus and novolimus DES.

Keywords: angioplasty; bioabsorbable polymer; drug-eluting stent; meta-analysis; stent.

Publication types

  • Meta-Analysis
  • Review

MeSH terms

  • Absorbable Implants*
  • Chi-Square Distribution
  • Coronary Artery Disease / diagnosis
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Coronary Restenosis / etiology
  • Coronary Restenosis / mortality
  • Coronary Thrombosis / etiology
  • Coronary Thrombosis / mortality
  • Drug-Eluting Stents*
  • Humans
  • Myocardial Infarction / etiology
  • Myocardial Infarction / mortality
  • Odds Ratio
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Polymers*
  • Prosthesis Design
  • Risk Factors
  • Time Factors
  • Treatment Outcome

Substances

  • Polymers