Preliminary results: prospective clinical study to assess image-based iterative reconstruction for abdominal computed tomography acquired at 2 radiation dose levels

J Comput Assist Tomogr. 2014 Jan-Feb;38(1):117-22. doi: 10.1097/RCT.0b013e3182a17629.

Abstract

Objective: The objective of this study was to compare image quality for abdominal computed tomographic (CT) images acquired at 200 and 50 mA s and reconstructed with image-based iterative reconstruction.

Materials and methods: In this institutional review board-approved prospective study, 22 patients (mean [SD] age, 64.3 [14.4] years; male-female ratio, 12:10) gave informed consent for acquisition of additional abdominal CT images on 64-slice multi-detector CT (MDCT) (Siemens Definition Flash). Standard-dose images were acquired at 200 quality reference mA s, whereas low-dose images were acquired at 50 mA s (all series: 120 kV; 5-mm section thickness; pitch, 0.9:1). The low-dose images were reconstructed with a nonlinear 3-dimensional iterative image reconstruction (3D-IIR) (SafeCT; MedicVision, Tirat Carmel, Israel) (4 settings, namely, A1, A2, A3, and A4) and were assessed by 3 abdominal radiologists for lesion detection, image noise, and visibility of small structures. CATPHAN 500 was scanned at the respective doses to obtain noise spectral density and modulation transfer function.

Results: Subjective image noise was unacceptable at 50-mA s filtered back projection and improved to average in 50-mA s A1 and minimal or no noise in 50-mA s A4. However, the visibility of small structures was similar to standard-dose filtered back projection images on 50-mA s A2. Objective image noise was reduced to 66% for the 50-mA s 3D-IIR images (9.08 [2.3]/26.75 [6.8]). The modulation transfer function curve demonstrated resolution improvement in the low-dose images with the 3D-IIR technique, whereas the noise spectral density curve confirmed noise suppression in the 50-mA s 3D-IIR images.

Conclusions: Three-dimensional iterative image reconstruction helps to lower image noise without affecting the visibility of small structures at "moderate" settings. Diagnostically acceptable abdominal CT examinations can be acquired at 75% lower-radiation dose with the help of the image-based iterative reconstruction technique.

Publication types

  • Comparative Study

MeSH terms

  • Contrast Media
  • Female
  • Humans
  • Imaging, Three-Dimensional
  • Iopamidol
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Prospective Studies
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Radiography, Abdominal
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media
  • Iopamidol