Field tested milliliter-scale blood filtration device for point-of-care applications

Biomicrofluidics. 2013 Aug 5;7(4):44111. doi: 10.1063/1.4817792. eCollection 2013.

Abstract

In this paper, we present a low cost and equipment-free blood filtration device capable of producing plasma from blood samples with mL-scale capacity and demonstrate its clinical application for hepatitis B diagnosis. We report the results of in-field testing of the device with 0.8-1 ml of undiluted, anticoagulated human whole blood samples from patients at the National Hospital for Tropical Diseases in Hanoi, Vietnam. Blood cell counts demonstrate that the device is capable of filtering out 99.9% of red and 96.9% of white blood cells, and the plasma collected from the device contains lower red blood cell counts than plasma obtained from a centrifuge. Biochemistry and immunology testing establish the suitability of the device as a sample preparation unit for testing alanine transaminase (ALT), aspartate transaminase (AST), urea, hepatitis B "e" antigen (HBeAg), hepatitis B "e" antibody (HBe Ab), and hepatitis B surface antibody (HBs Ab). The device provides a simple and practical front-end sample processing method for point-of-care microfluidic diagnostics, enabling sufficient volumes for multiplexed downstream tests.