Abstract

The research work presented in this paper focuses on qualitative tissue differentiation by monitoring the intensity ratios of atomic emissions using 'Laser Induced Breakdown Spectroscopy' (LIBS) on the plasma plume created during laser tissue ablation. The background of this study is to establish a real time feedback control mechanism for clinical laser surgery systems during the laser ablation process. Ex-vivo domestic pig tissue samples (muscle, fat, nerve and skin) were used in this experiment. Atomic emission intensity ratios were analyzed to find a characteristic spectral line for each tissue. The results showed characteristic elemental emission intensity ratios for the respective tissues. The spectral lines and intensity ratios of these specific elements varied among the different tissue types. The main goal of this study is to qualitatively and precisely identify different tissue types for tissue specific laser surgery.

Keywords: Laser induced breakdown spectroscopy; chemical composition; chemical elements; intensity ratio; plasma; tissue type.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Feedback*
  • Laser Therapy*
  • Organ Specificity
  • Spectrum Analysis*
  • Swine