Optimization of coronary attenuation in coronary computed tomography angiography using diluted contrast material

Circ J. 2014;78(3):662-70. doi: 10.1253/circj.cj-13-1101. Epub 2013 Dec 21.

Abstract

Background: The purpose of this study was to evaluate a personalized protocol with diluted contrast material (CM) for coronary computed tomography angiography (CTA).

Methods and results: One hundred patients with suspected coronary artery disease underwent retrospective electrocardiogram-gated coronary CTA on a 256-slice multidetector-row CT scanner. In the diluted CM protocol (n=50), the optimal scan timing and CM dilution rate were determined by the timing bolus scan, with 20% CM dilution (5ml/s during 10s) being considered suitable to achieve the target arterial attenuation of 350 Hounsfield units (HU). In the body weight (BW)-adjusted protocol (n=50, 222mg iodine/kg), only the optimal scan timing was determined by the timing bolus scan. The injection rate and volume in the timing bolus scan and real scan were identical between the 2 protocols. We compared the means and variations in coronary attenuation between the 2 protocols. Coronary attenuation (mean±SD) in the diluted CM and BW-adjusted protocols was 346.1±23.9 HU and 298.8±45.2 HU, respectively. The diluted CM protocol provided significantly higher coronary attenuation and lower variance than did the BW-adjusted protocol (P<0.05, in each).

Conclusions: The diluted CM protocol facilitates more uniform attenuation on coronary CTA in comparison with the BW-adjusted protocol.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Body Weight
  • Contrast Media / administration & dosage*
  • Coronary Angiography / methods*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media