Influence of measurement depth on the stiffness assessment of healthy liver with real-time shear wave elastography

Ultrasound Med Biol. 2014 Mar;40(3):461-9. doi: 10.1016/j.ultrasmedbio.2013.10.021. Epub 2013 Dec 19.

Abstract

The purpose of this study was to determine the measurement depth range within which liver stiffness can be reliably assessed using real-time shear wave elastography (SWE) technology. Measurements were performed on phantoms and healthy volunteers. In the first group of patients, measurements were performed at depths of 2-8 cm from the probe surface. In the second group of patients, measurements were conducted 0-7 cm below the liver capsule. Success rate of measurements (SRoM), success rate of patients (SRoS) and coefficients of variation (CVs) of repeated measurements were compared. The SRoMs at 3-7 cm and the CVs at 2-5 cm from the probe surface were significantly higher and lower than those at other depths (p < 0.001), respectively. SRoS was zero 0-1 cm below the liver capsule. Furthermore, the features of 2-D stiffness mapping images were also found to change with depth. According to our results, the depth range for the most reliable liver stiffness assessment using SWE should be 3-5 cm from the probe surface and simultaneously 1-2 cm below the liver capsule.

Keywords: Liver capsule; Liver stiffness; Measurement depth; Real-time shear wave elastography; Young's modulus.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Algorithms*
  • Elastic Modulus / physiology
  • Elasticity Imaging Techniques / instrumentation
  • Elasticity Imaging Techniques / methods*
  • Female
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods*
  • Liver / diagnostic imaging*
  • Liver / physiology*
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Reference Values
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Shear Strength