Development of stable and reproducible biosensors based on electrochemical impedance spectroscopy: three-electrode versus two-electrode setup

Biosens Bioelectron. 2014 May 15:55:1-6. doi: 10.1016/j.bios.2013.11.067. Epub 2013 Dec 4.

Abstract

This work focuses on the development of electrochemical impedance biosensors based on capacitance readout, for the detection of biomolecules in small sample volumes. We performed electrochemical impedance spectroscopy (EIS) measurements of DNA hybridization in electrochemical cells with microfabricated gold electrodes. The time stability of the device was tested in two different configurations: two microelectrodes in a microfluidic channel; two microelectrodes plus a reference electrode in an electrochemical cell. Our results demonstrate that the three-electrode setup is more stable, more reproducible, and suitable for real-time measurements. In the last part of the work we perform a test study of DNA hybridization in real time, and we show that the three-electrode configuration can measure the process in situ and in real time.

Keywords: DNA hybridization; DNA self-assembling; EIS; Real-time biosensors.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / instrumentation*
  • DNA / analysis
  • DNA / chemistry*
  • DNA / genetics*
  • Dielectric Spectroscopy / instrumentation*
  • Electrodes*
  • Equipment Design
  • Equipment Failure Analysis
  • In Situ Hybridization / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity