Performance of the new HIV-1/2 diagnostic algorithm in Florida's public health testing population: a review of the first five months of utilization

J Clin Virol. 2013 Dec:58 Suppl 1:e29-33. doi: 10.1016/j.jcv.2013.08.016. Epub 2013 Aug 27.

Abstract

Objective: The Centers for Disease Control and Prevention and the Association of Public Health Laboratories have proposed a new HIV-1/2 Diagnostic Algorithm: a fourth-generation HIV-1/2 Ag/Ab immunoassay (IA) followed, when repeatedly reactive, by an HIV-1/HIV-2 antibody differentiation test, and if that is non-reactive, HIV-1 nucleic acid amplification testing (NAT). The objective of the study was to evaluate performance of the new algorithm after five months of utilization in our high volume, high HIV-1 seroprevalence public health population.

Methods: Algorithm sensitivity and specificity was evaluated on 51,953 prospective serum or plasma specimens from individuals self-referring for HIV serostatus determination. Specimens were tested on the day of receipt or maintained at 4°C until the next testing opportunity. If the initial HIV-1/2 Ag/Ab IA (Abbott Combo) was nonreactive, a negative lab interpretation report would follow. If the initial IA was reactive, repeat screening in duplicate was immediately performed. Repeatedly reactive specimens were tested with an HIV-1/HIV-2 differentiation assay (Multispot [MS] HIV-1/HIV-2 Rapid Test) on the same or next workday. If the Abbott Combo-MS assays were discordant, HIV-1 NAT (APTIMA(®) HIV-1 RNA) was performed. In addition to the algorithm performance, we also evaluated the laboratory "specimen receipt to reporting" turnaround time (TAT).

Results: The sensitivity and specificity of the new HIV Diagnostic Algorithm with serum and plasma specimens over the initial 5 month period was 100% (922/922) and 99.99% (51,030/51,031), respectively. Two algorithm-defined acute HIV-1 infections (AHI) were detected. In addition only 3 of the 992 MS secondary tests performed were interpreted as HIV-1 Indeterminate (HIV-1 recombinant gp41 reactivity only). Of these, 2 were HIV-1 NAT reactive, defined in-house as an early HIV infection (EHI) and one was HIV-1 NAT nonreactive, indicating a false positive initial screening result. Laboratory TAT for reporting concordant reactive Abbott Combo-MS results in ≤ 2 workdays was 96%, compared to 22% for reporting concordant reactive 3rd generation IA-Western blot results.

Conclusions: In our public health testing population, results from the new HIV Diagnostic Algorithm exceeded those of the 3rd generation IA-WB algorithm with respect to HIV-1 sensitivity. The identification of two algorithm-defined AHIs provided the opportunity to inform these individuals of their HIV status and link them to medical care earlier than the scheduled posttest counseling appointment.

Keywords: HIV; New HIV Diagnostic Algorithm; Public health.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Diagnostic Tests, Routine / methods*
  • Florida
  • HIV Antibodies / blood*
  • HIV Antigens / blood*
  • HIV Infections / diagnosis*
  • HIV Infections / virology
  • HIV-1 / classification*
  • HIV-1 / genetics
  • HIV-1 / immunology
  • HIV-2 / classification*
  • HIV-2 / genetics
  • HIV-2 / immunology
  • Humans
  • Immunoassay / methods
  • Nucleic Acid Amplification Techniques / methods
  • RNA, Viral / blood*
  • Sensitivity and Specificity
  • Virology / methods

Substances

  • HIV Antibodies
  • HIV Antigens
  • RNA, Viral