An analysis of the gradient-induced electric fields and current densities in human models when situated in a hybrid MRI-LINAC system

Phys Med Biol. 2014 Jan 6;59(1):233-45. doi: 10.1088/0031-9155/59/1/233. Epub 2013 Dec 13.

Abstract

MRI-LINAC is a new image-guided radiotherapy treatment system that combines magnetic resonance imaging (MRI) with a linear accelerator (LINAC) in a single unit. One drawback is that the pulsing of the split gradient coils of the system induces an electric field and currents in the patient which need to be predicted and evaluated for patient safety. In this novel numerical study the in situ electric fields and associated current densities were evaluated inside tissue-accurate male and female human voxel models when a number of different split-geometry gradient coils were operated. The body models were located in the MRI-LINAC system along the axial and radial directions in three different body positions. Each model had a region of interest (ROI) suitable for image-guided radiotherapy. The simulation results show that the amplitudes and distributions of the field and current density induced by different split x-gradient coils were similar with one another in the ROI of the body model, but varied outside of the region. The fields and current densities induced by a split classic coil with the surface unconnected showed the largest deviation from those given by the conventional non-split coils. Another finding indicated that the distributions of the peak current densities varied when the body position, orientation or gender changed, while the peak electric fields mainly occurred in the skin and fat tissues.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electric Conductivity*
  • Female
  • Humans
  • Magnetic Resonance Imaging / adverse effects
  • Magnetic Resonance Imaging / instrumentation*
  • Male
  • Models, Anatomic
  • Particle Accelerators*
  • Patient Safety